Dental Delivery Unit Operation Manual
REV-A
Page 18 of 45
Please ensure the place has a separate and stable power source supply.
Please ensure the place has the sewage draining tube to connect.
Please ensure the place has air supply.
Please ensure the palace has water supply.
The temperature of environment:5-40
℃
.
The humidity of environment: ≤85%
Atmosphere pressure:860Hpa
~
1060Hpa.
4.2 Power Connection
The dental unit should be connected with a stable and separate power source
Please confirm the connection power is complied with specification of nameplate at the
back of dental unit.
CAUTION
DO NOT BEND THE POWER WIRE TO AVOID DAMAGE OF POWER WIRE
DO NOT PUT ANY HEAVY STUFF ON THE POWER WIRE TO AVOID DAMAGE
OF POWER WIRE
DO NOT USE OTHER POWER WIRE TO AVOID DAMAGE OF DENTAL UNIT
DO NOT TRY TO ADD THE POWER WIRE TO AVOID ACCIDENT AND RISK
WRONG
WRONG
RIGHT
Set Chair and Junction box
CAUTION
THE GRADIENT SHOULD BE LESS THAN 3°. PLEASE ENSURE THE
GROUND IS HORIZONTAL, OTHERWISE IT WILL EFFECT OPERATION AND
PERFORMANCE, THIS MIGHT EVEN CAUSE ACCIDENTS
Before setting the Junction box the please use a hose to exhaust dirty water or air inside the
tube, for cleaning the impurity and dirt.