1. OPERATOR MODE
Status bar
Under the window with video signal is status bar, which display informa on about state of system (Live Stream,
measurement, etc.) on the le , in the middle, according to se ngs in tab
App. Se ngs
, informa on about
current zoom of video signal or temperature of camera or current value of frame per second and resolu on
of camera. On the right is indicator
Time to Calib
. which counts down the me to the next automa c image
compensa on. During automa c image compensa on is video signal paused for moment.
Temperature range and pale e of thermographic image
To the right of window with video signal is temperature pale e of thermographic image. All temperatures be-
tween maximal and minimal limits (right of pale e) are con nuously alloca ng color. From pale e can be easy
read what temperature has specific area in the thermographic image.
Bu on Full Screen
This bu on will launch the Full Screen Mode, which is described in Chapter 3. This bu on can be pressed only
when is ac ve streaming image from the camera and running inspec on. Because this applica on is only display
and it is not possible to set anything or launch.
Table of defini on of measurement area (ROI)
To the right of pale e is table containing se ngs for each defined measurement area. Each area in image (ROI)
corresponds to one row in the table. Table contains following columns:
ROI
– Numeric iden fier which assigns row from table to specific ROI in image.
Trigger
– Indicate which trigger is assigned to each ROI.
Sta s c
– Sta s c which is evaluated in ROI. Sta s c is calculated for all pixels in ROI.
Current Value
– Current value of sta s c in ROI.
Criterion
– Criterion which is in sta s c checked during the measurement.
Last Result
– Value of criterion a er last measurement.
Operator
– Comparison operator with which is evaluated Last Result against
Criterion Value
during mea-
surement.
Criterion Value
– Set reference value of criterion.
Outputs 1-5
– In these columns are marked by cross linkage row to digital output of system. Etc. when all
condi ons are met, marked output is ac ve.
The table in Figure 1.5 is primary used to representa on current se ng criterion, reference value, performing
operator and last measured value. If the last measured value of criterion met the condi on of comparison op-
erator and reference value, whole row is colored green a er each measurement. Otherwise, row is colored
red.
In the last row of table on the le is displayed se ng behavior of digital output. Are used two op on
All
and
One
. First op on
All
indicate that for ac vate output in the column (Outputs 1-5) have to be met condi ons in
all rows which have the cross in corresponding column (or row a er the measurement is green). Second op on
One
indicate that for ac ve output in the column have to be met condi ons in at least of one rows which have
the cross in corresponding column.
Headers of individual column of outputs labeled 1-5 is lighted up yellow when the digital output is ac ve.