2. ADVANCED SETTINGS AND ADMINISTRATOR MODE
2.1.4 Temperature Calibra on
Figure 2.6 – Se ngs Temperature Calibra on
Temperature calibra on may be 1 to 3 points, according to the number of filled parameters.
Real Calibra on
Temperature
is the actual value that indicates the camera. It can be set in two ways depending on the se ng
switch
Set in Image / Numerically
. If it is set to
Click
, you can click on the
Set Loca on
and choose the desired
loca on in the thermal image and value at this loca on is updated every 2 seconds and temperature calibra on
is recalculated. If the switch is set to
Num
, is required value
Real Calibra on Temperature
enter on the keyboard.
Black Body Calibra on
is the value that the camera should indicate in the point, this value is entered on the
keyboard. Calibra on is ac vated or deac vated by clicking tot check-box
Load Calibra on
. To calibrate individual
cameras is necessary to be selected the relevant camera in field the
Select Camera
. It is possible to calibrate each
camera. When selec ng
All Cameras
calibra on cannot be set but shows the calibrated image that has been set
for each camera.
2.1.5 Set Differen al Image
Figure 2.7 – Se ng Differen al Image