2. ADVANCED SETTINGS AND ADMINISTRATOR MODE
ment, but not before the expiry me set in the numeric field to the right (this restric on applies to all
compensa on associated with measurements).
Synchronous a er a Measurement
t – The compensa on is performed a er comple on of measurement.
Synchronous before and a er a Measurement
– The compensa on is performed before start and a er
comple on of measurement.
Automa c compensa on of the camera image is performed automa cally a er the me interval set in
Syn-
chronous a er me
regardless of the above se ng. For example, if the implementa on of compensa on is set
at the beginning of each measurement, and the measurement takes longer than the set interval, compensa on
is performed automa cally a er the set interval (and measurement con nues).
Select Camera
Field
Image freezing me during calibra on
sets the me that the video is paused when automa c compensa on
is performed. This me depends on the specific type of camera it is generally in the range of 500 ms - 1500 ms.
Select Camera
Sets parameters for automa c reconnec on camera when it is disconnected or an error in the displayed data.
Reconnect Interval
– The operator is alerted to the disconnected camera in case that during this me
connec vity error occurs two mes.
Time to Reconnect
– A me period a er which the program automa cally searches disconnected the cam-
era. A er this me, if the camera is not found, alerts the operator.