![WORKSWELL THERMOINSPECTOR Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/workswell/thermoinspector/thermoinspector_user-manual_997409014.webp)
2. ADVANCED SETTINGS AND ADMINISTRATOR MODE
2.1.2 Image Setup
Figure 2.4 – Se ngs Image Setup
Object Emissivity
Sets emissivity measured objects (0-1). In the case of more frequent compara ve (qualita ve) measurement
(e.g. comparing temperature increase, finding hot spots) can leave the preset emissivity value and not to change
it during the measurement. If required precise measurements, select the emissivity value for the monitoring
material in the table of emissivity values (see A achments manual).
Windows Transmission
Sets the rela ve transmission of the protec ve window in front of the camera. If there is no protec ve window
used, must be set to 1. The transmission of the protec ve germanium window is at normal opera ng tempera-
ture equal to 0.98, but decreases with rising temperature. If the ambient temperature is during the measurement
higher than 50 ° C, reduce the transmission to 0.85. Accurate determina on of transmission is carried out using
a calibra on black-body. For qualita ve measurements at a constant ambient temperature it is not necessary to
change the transmission.
Humidity
Sets the humidity of environment (0-100%).