2. ADVANCED SETTINGS AND ADMINISTRATOR MODE
dialog (see Figure 2.21).
Figure 2.21 – Edi ng dialog for ROI
Descrip on
– A short descrip on, which appears in the window with the video signal under the iden fier
of the ROI. If the field is empty, it does not appear any label.
Descrip on Color
– The color of the label text and iden fier ROI.
Background Color
– The background color of text label and iden fier ROI. The color selector is possible to
choose a transparent background text (upper right field with the emblem T)
Graph Visibility
– If the box is checked, the sta s cs of the ROI will be plo ed in all charts in the applica on.
Display Maximum Temperature
– If the box is checked, the in the window with the video signal is shown
the current maximum temperature and its loca on.
Display Minimum Temperature
– If the box is checked, the in the window with the video signal is shown
the current minimum temperature and its loca on.
Enable Min Max Temperature over TCP/IP
– To enable/disable sending the values of minimum and maxi-
mum temperatures over the TCP/IP protocol.
Enable Envelope for Evalua on
- If the box is checked, ROI will be evaluated according created envelope.
Switch Global/Local
– Sets whether is required for a desired ROI used to determine the temperature of the
Global emissivity, which is set in the tab
Image Source Setup
in the menu
Image Setup
in the field
Object
Emissivity
. Otherwise you can use local emissivity, which can be set for each ROI independently.