Wolfgang Warmbier GmbH & Co. KG
25
11
Characteristic Values
METRISO 3000
1
With the included, shielded, high-impedance measurement cables
2
DC measuring voltage U
M
5%
3
See figure 1 on page 26
4
stable as from 200 k
U
M
5
U
M
10 V: +7 Digit
Reference Conditions
Reference
temperature
+ 23
C
3 K
Relative humidity
40 ... 75%
Measured quantity
frequency
45 Hz ... 65 Hz
Measured quantity
waveform
Sine, deviation between TRMS and recti-
fied value < 1%
Battery voltage
9.5 V
0.1 V
Test resistance
10 M
1%
Electrical Safety
Protection class
II per IEC/EN 61010-1/VDE 0411-1
Pollution degree
2
Measuring category
CAT II, 600 V
Fuses
Fuse link
FF315mA/1000V, effective in all resis-
tance measuring ranges, 1 additional
replacement fuse in the battery compart-
ment
Electronic fuse
For protecting resistance measurement R
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2006, class B
Interference immunity EN 61326-1:2006
Power Supply
Batteries
8 ea. 1.5 V mignon cell (8 ea. size AA)
(alkaline manganese per IEC LR14)
or 8 rechargeable NiMH batteries (must be
recharged externally)
Z502R charger
Broad-range charger with barrel connec-
tor, input: 100 to 240 V AC,
output: 16.5 V DC, 1 A (Mascot)
Nominal range of use 8.5 ... 12 V
Battery test
Battery capacity display with battery sym-
bol in 4 segments:
.
Querying of momentary battery voltage via
menu function.
Battery saver circuit
Automatic shutdown of display illumination
after 10 to 30 seconds (after the last time
the rotary switch is actuated) can be
selected in the setup menus (see page 8).
The test instrument is automatically
switched to the
standby mode
if the mea-
sured value remains unchanged for approx-
imately 15 minutes, and if none of the
controls are activated during this time.
The instrument is switched
off automatically
if the measured value remains unchanged
for a long period of time, and if none of the
keys or the rotary switch have been acti-
vated during on-time (specified in seconds).
Service life
3000 measurements with a set of
rechargeable batteries and a measuring
time of 5 seconds from one measurement
until automatic shutdown of the measuring
procedure.
Safety shutdown
If supply voltage is too low (U < 8 V), the
instrument is switched off, or cannot be
switched on.
Recharging socket
Installed rechargeable batteries can be
recharged directly by connecting a charger
to the recharging socket:
M
PRO
M
XTRA
G1000+ charger (Z502R)
Charging time
Approx. 2 hours *
* Maximum charging time with fully depleted rechargeable batteries.
A timer in the charger limits charging time to no more than 4 hours.
Displays
Digital Display
Multiple display with dot matrix,
128 x 128 pixels,
backlit (transflective),
dimensions: 65 x 65 mm
Limit LED
LED lights up red to indicate an exceeded
limit value
LED lights up green to indicate adherence
to the limit value
LED
LED lights up red to indicate the presence
of interference voltage (when instrument is
switched off) or high test voltage at the
measuring sockets during resistance mea-
surement (R
ESD
)
Meas.
Qty.
U
M
2
Range
Measuring Range
Reso-
lution
Intrinsic Error
1
Measuring
Uncertainty
Overload Capacity
R
ESD
10 V
3
10 k
1.0 k
... 9.99 k
0.01 k
(5% rdg. + 10 d)
(7% rdg. + 10d)
500 V AC/DC TRMS
100 V
3
500 V
3
100 k
10.0 k
... 99.9 k
0.1 k
(5% rdg. + 3 d)
5
(7% rdg. + 3 d)
1 M
4
100 k
... 999 k
1 k
(5% rdg. + 3 d)
(7% rdg. + 3 d)
10 M
1.00 M
... 9.99 M
10 k
(5% rdg. + 3 d)
(7% rdg. + 3 d)
100 M
10.0 M
... 99.9 M
100 k
(5% rdg. + 3 d)
(7% rdg. + 3 d)
1 G
100 M
... 999 M
1 M
(5% rdg. + 3 d)
(7% rdg. + 3 d)
10 G
1.00 G
... 9.99 G
10 M
(5% rdg. + 3 d)
(10% rdg. + 3 d)
100 G
10.0 G
... 99.9 G
100 M
(8% rdg. + 3 d)
(10% rdg. + 3 d)
1 T
100 G
... 999 G
1 G
(25% rdg. + 5 d)
±(50% rdg. + 20 d)
U
AC/DC
100 V
10.0 V ... 99.9 V
0.1 V
(2.5% rdg. + 3 d)
(5% rdg. + 3 d)
500 V AC/DC TRMS
500 V
100 V ... 499 V
1 V
R
Display range as of
01.0
100
1.0 ... 99.9
0.1
(2.5% rdg. + 3 d)
(5% rdg. + 3 d)
500 V AC/DC TRMS
1 k
100 ... 999
1
10 k
1.00 ... 9.99 k
10
!