Significance of Individual Parameters
Test Instrument On-Time
The period of time after which the test instrument is automatically
shut off can be selected here. This selection has a considerable
influence on the service life and the charging status of the batter-
ies.
Note
As long as the T/F sensor is connected to the RS 232
interface, automatic shutdown of the ESD test instrument
is disabled.
On-Time for LCD Illumination
The period of time after which LCD illumination is automatically
shut off can be selected here. This selection has a considerable
influence on the service life and the charging status of the batter-
ies.
Submenu: Battery Level Query
Battery voltage U
BAT
(Uo) can be queried here without subjecting
the battery to a load.
Note
Measuring Sequence
If battery voltage drops to below 8.0 V
during the course of a measuring
sequence, this is only indicated by
means of a pop-up window. Measured
values are invalid. Measurement results cannot be saved
to memory.
➭
Press
ESC
in order to return to the main menu.
Attention!
!
Data are lost when the
language
is changed, or if the instrument is
reset to its
default values
!
Back up your measurement
data to a PC with the help of
ETC software before press-
ing the respective key.
The prompt window shown at
the right asks you to confirm
deletion.
User Interface Language (CULTURE = LANGUAGE)
➭
Select the desired language with the appropriate country
code and acknowledge your selection by pressing the
✓
key.
Default Settings (GOME SETTING)
The test instrument is returned to its original default settings when
this key is activated.
Firmware Revision and Calibration Information (example)
➭
Press any key in order to return to the main menu.