Wolfgang Warmbier GmbH & Co. KG
13
5
Measuring Direct and Alternating Voltage
You can measure direct voltage, as well as sinusoidal alternating
voltage, with frequencies ranging from 45 to 65 Hz with this test
instrument.
Select Measuring Function
➭
Select the
V
measuring function with the rotary switch.
Connection
➭
Connect the measurement cables to the
+
and
COM
jacks.
➭
Contact the measuring point with both test probes.
Set Voltage Type Parameter (waveform)
Measurement
The measured value is displayed directly (without pressing the
START
key) – in analog format at the bar graph and in digital format
at the matrix display.
➭
After completing the measurement, switch the instrument off
by turning the rotary switch to the
OFF
position.
The
ESC
,
START
and
CONTIN
keys have no function in this case.
Note
Input impedance for the voltage measuring range is
10 M
.
V
L1
L2
N
PE
L3
R
B
R
E
Max
.
600 V
U
+
COM
+
COM SHIELD
1
2
2
3
Direct voltage
Alternating voltage
Pulsating voltage (DC + AC TRMS)