Fig. 55: Configuration of the logger
measurement
•
Enter the start date and time
•
Enter the measurement interval, the interval
duration in the sample in the figure on the left
is 10 seconds
•
Enter the number of measurements (up to
9,999 measurements are possible).
NOTE!
The analyzer calculates the date and time of
completion of the logger measurement for infor-
mation purposes.
•
Select Purging 'On', to activate the fresh air
pump in a selected interval. (only available
when interval duration is < 8 min).
In this case the sensors will be purged with fresh
air in the selected purging interval.
NOTE!
During a purging cycle the analyzer will not log
measurements. A purging cycle needs about 3
minutes to complete.
Long time measurements with
the USB Peltier cooler
For continuous long term measurements with
interval < 8 min it is recommended to use the
USB Peltier cooler. The peltier cooler will remove
condensate from the sample gas. During opera-
tion the condensate trap of the cooler needs to be
emptied.