27
WIKA operating instructions SF
6
gas filling cart, model GFU08
14033074.01 11/2018 EN/DE/FR/ES
EN
Step
Display
Description
6
hirES
Non-calibratable display unit.
or
LEGAL
Calibratable display unit.
7
Display unit performs countdown for self-checking.
7.2.1 Switching off
Keep the C button pressed until the message “
-off-
” is shown in the display.
In order to the switch the display unit off and then on again, keep the C button pressed
for 8 seconds. The display unit switches off and then automatically switches back on.
7.3 Zeroing the scales
If the gross weight lies within ± 2 % of the nominal load, it is possible to set the display
unit to zero by pressing the ZERO button. The display no longer indicates any weight
and the power-on lamp lights up, and symbols are shown on the display.
7.4 Function of the MODE button
Operating
modes
But-
ton
Effect
Button Effect
Button
STANDARD
MODE
Change units
between kg and lb
NET/GROSS
MODE
Switching between
gross and net
weight
SETPOINT
ENTER With model GFU-08
SF
6
gas filling cart,
without function
IN/OUT
MODE
Input weight
ENTER Output weight
ALIBI MEMORY
MODE
Input of recall
number
ENTER Input of ID number
ENTER
TOLERANCE
CHECK
MODE
Input of target
weight
ENTER Input of minus
tolerance weight
>>>>>
ENTER Input of plus
tolerance weight
ENTER Input of minimum
weight
ENTER
PERCENT
MODE
Input of % value
ENTER MODE button:
a) Changes % to
weight
MODE button 2 sec.:
b) Piece
7. Display unit