8. Instrument messages during operation
Display Description
2ERO
The scales are trying to effect a zero setting
AL.Err
The alibi memory is selected, but on starting the scales, no alibi memory is connec-
ted or the data communication between the display and the keypad is faulty.
Er I.b.H
A function is linked to input X (from 1 to 4) and is not
“
present
”
. See SETUP area
under
“
input5
”.
Er.r.b.H
In a switch point function, the relay X (from 1 to 4) has been set and is not “present”.
See SETUP area under
“
output
”
.
busy
Printing is in progress (serial connection PRN is busy) or the display unit is waiting to
be able to send a print to the PC.
unstAb
Printing attempt with an unstable weight.
un.oUEr
Printing attempt in over or underload condition, i.e. with a weight that is 9 scale divisi-
ons greater than the nominal load or 100 scale divisions under the gross zero.
LoW
The net weight is less than the minimum load required for printing or totalling, or less
than the minimum weight for transferring a data string by pushbutton.
no.0
..un5
Weight has not passed net 0 or was not unstable (depending on configuration).
ConU.
On legally calibrated instruments, an attempt was made to print in the standard
setting while the instrument was adjusting the weight unit.
no in
In the input/output function (activated as
“
in .out
”
), a second attempt was made to
acquire the input weight.
no out
In the input/output function (activated as
“
in .out
”
), a second attempt was made to
acquire the output weight.
no 1
In the input/output function (activated as
“
G.t.
”
or
“
1St.2nd
”
), a second attempt was
made to acquire the input weight.
no 2
In the input/output function (activated as
“
G.t.
”
or
“
1St.2nd
”
), a second attempt was
made to acquire the output weight.
Er.Not
Unstable weight.
Error
In counting mode, the reference determination was not or not properly performed, or
a larger reference quantity is required. During the comparison procedure, the load
cell with which the comparison is performed does not correspond to the rated load or
values have not been entered or have been entered in the wrong order.
E9
..Err
Is displayed when the execution of the comparison was not possible.
StorE
Is displayed when data has been stored in the permanent memory of the instrument
(switch points, tare, updated record, etc.)
Err.Clk
Problem with the date/time cards. Setting of
“
F.NodE
”
>>
“
Clock
”
in SETUP.
SEt.Clk
Date/Time not set. Setting of
“
F.NodE
”
>>
“
Clock
”
in SETUP.
PrEC.
Is displayed when attempting to perform the zero calibration without first confirming
the number of calibration points.
8. Instrument messages during operation