46
6 Set up the sensor with the display and adjustment module
WIKA Operating Instructions - Differential pressure transmitter DPT-20
6.5.2 Display
This menu item enables the setting of the requested national lan
-
guage.
The following languages are available:
•
German
•
English
•
French
•
Spanish
•
Russian
•
Italian
•
Dutch
•
Portuguese
•
Japanese
•
Chinese
•
Polish
•
Czech
•
Turkish
In delivery status, the DPT-20 is set to English.
In this menu item, you define which measured value is displayed.
The default setting for the displayed value is "
Differential pressure
".
In this menu item you define the number of decimal positions with
which the measured value is displayed.
The default setting for the display format is "
Automatic
".
The display and adjustment module has a backlight for the display.
In this menu item you can switch on the lighting. You can find the
required operating voltage in chapter "
Technical data
".
In delivery status, the lighting is switched on.
Language
Display value 1 and 2 -
bus systems
Display format 1 and 2
Backlight