15
WIKA operating instructions, model CPH6300
EN
14043054.02 08/2017 EN/DE
Menu
Parameters
Values
Meaning
“MENU”
▶
▲ or ▼
P
.
oFF
Auto power-off: Automatic switching off of the instrument
1 ... 120
Auto power-off delay in minutes. If no button is
pressed and there is no data transfer via the
serial interface, the instrument will switch itself off
after this interval.
oFF
Auto power-off function inactive (continuous
operation)
Lite
Backlighting (factory setting: 5 s)
oFF
No illumination
5 ... 120
Illumination automatically switches off after
5 ... 120 s.
on
Illumination does not switch off
SEt
CALC
Set calculation: Calculated displays (derived from sensor 1)
1)
CALC
Selection of the calculation function
oFF
No calculated display
dP:dt
Rate of change of pressure
SPEd
Airspeed through the orifice plate/pitot tube
FLo
Flow / Air flow
“dP:dt”: Settings
base
Time unit for the rate of change of the pressure
PEr.S
Change in pressure per second, e.g. mbar/s
PEr.n
Change in pressure per minute, e.g. mbar/min
PEr.h
Change in pressure per hour, e.g. mbar/h
t
.
int
Test duration, for which the pressure change display is
calculated
00:01 ... 1:00
Duration in [Minutes:Seconds], Factory setting:
0:01 = 1 s
1) Menu cannot be selected if there is data in the logger memory. If this is to be changed, the data must first be deleted (see chapter
5.5 “Data logger”).
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding
sensor on connection 2 then the settings are adopted.
3) Only for 2-channel version
5. Commissioning, operation