Display
Meaning
Remedy
Err
.
1
Measured value above allowable
range
Check: Is the pressure over the
permissible one? Measuring range of
the sensor?
⇒
Measured value too high!
Sensor defect
Send in for repair
Err
.
2
Measured value below allowable
range
Check: Is the pressure under the
permissible measuring range of the
sensor?
⇒
Measured value too low!
Sensor defect
Send in for repair
Err
.
3
Scale range exceeded
Check: Is the value above 19999?
⇒
Measured value too high!
Err
.
4
Under the scale range
Check: Is the value below -19999
(Tare?)
⇒
Measured value too low!
Er
.
11
Measured value could not be
calculated
Choose different unit
Calculation range exceeded
Choose different unit
Err
.
7
System error
Send in for repair
----
Sensor not present / recognised
Plug in inserted sensor once more,
with running recording, stop the
logger and re-start it
Display value cannot be calculated
Plug in appropriate sensor
combination
6. Faults / 7. Maintenance, cleaning and recalibration
7. Maintenance, cleaning and recalibration
Personnel:
Skilled personnel
Protective equipment:
Protective gloves and safety goggles
Tools:
Cross-head screwdriver, spanner or torque spanner
For contact details see chapter 1 “General information” or the back page of
the operating instructions.
7.1 Maintenance
The model CPH6300 hand-held pressure indicator and the associated CPT6200
reference pressure sensors are maintenance-free.
Repairs must only be carried out by the manufacturer.