26
WIKA operating instructions, model CPH6300
EN
14043054.02 08/2017 EN/DE
5.5.1.3 Clear manual recording
If the data has already been saved, this can be cleared via the
[STORE/QUIT]
button:
1. Press the
[STORE/QUIT]
button for
2 seconds.
⇒
Accesses the “Delete” function.
2. Press the
[
▲
]
or
[
▼
]
button.
⇒
Change the selection.
CLr no
Do not delete (cancel the process).
CLr ALL
Delete all data sets.
CLr LASt
Clear the last data set.
3. Press
[STORE/QUIT]
button.
⇒
Confirm the selection, end of the “Delete” function.
5.5.2 Automatic recording with adjustable cycle
If the logger function,
Fvnc CYCL
, is selected (see 5.3 “Configuration of the
instrument”), following the start of the logger, the measured value is automatically at the
interval of the set cycle time.
The logger cycle time is adjustable from 1 s ... 60 min.
At the moment of saving, a data set consists of the following values.
(1-channel version):
Storable data sets: 10,000
With slow measurements:
rAtE SLo
■
Measured value
■
CALC off
: Min.-Peak and Max.-Peak since last save
■
CALC
“active”: Estimated calculation and dummy values (see chapter 5.4.5 “Rate of
change of pressure (only for sensor connection 1)”)
With fast measurements:
rAtE FASt
and
rAtE P
.
dEt
■
Arithmetic average since last save
■
CALC off
: Min.-Peak and Max.-Peak since last save
■
CALC
“active”: Estimated calculation and dummy values (see chapter 5.4.5 “Rate of
change of pressure (only for sensor connection 1)”)
5. Commissioning, operation