Single Channel ECG Test System | User Manual
(continued)
design on a simple triangle pulse which may
explain small differences in test results.
50.102.16
Output display
No special guidance.
50.102.17
Tall T-wave
rejection
This test is intended to verify that the patient
monitor shows the correct heart rate in the
presence of a high T-wave. At some point, most
patient monitors will treat the high T-wave as
new QRS pulse, and hence double count the
heart rate (e.g. display 160bpm for an 80bpm
input). Note a T-wave of 1mV will actually
appear higher than the QRS with an amplitude
of 1mV due to the characteristics of Figure 119.
Note: In the future it is planned to update this table to the IEC 60601-2-27:2011
edition