Single Channel ECG Test System | User Manual
2.6
Description of Functional Groups
2.6.1
Main function (main waveform)
This group allows the operator to select the main waveform to be used in the test,
from the following:
Waveform
type
Description
Sample waveform
Sine
Basic sine wave, according to
the amplitude (mVpp) and
frequency (Hz or bpm)
Triangle
Basic triangle wave, according
to the amplitude (in mVpp)
and frequency (Hz or bpm)
Square
Basic square wave, according
to the amplitude (mVpp) and
frequency (Hz or bpm)
Rectangle
pulse
A rectangular pulse, according
to the amplitude setting, pulse
width (ms) and pulse
repetition rate (frequency, Hz
or bpm)
Triangle
pulse
A triangle pulse, according to
the amplitude setting, base
pulse width (ms) and pulse
repetition rate (frequency, Hz
or bpm)
Exponential Exponential waveform, used to
Hysteresis test, according to the
amplitude (mVpp) and
frequency (Hz or bpm)
ECG
Waveform according to IEC
60601-2-27, Figure 201.110
and 201.113, with adjustable
parameters for amplitude
(mVpp), frequency (Hz or
bpm), QRS duration and T
wave amplitude
Special
A range of stored waveforms
including ANSI/AAMI
waveforms, some selected CAL
waveforms, and loaded