SECG 4.0 | User Manual
2.1.4
First Time Using WhaleTeq Product
–
Microsoft .Net Framework 4.0
Installation
WhaleTeq software is developed by Microsoft .Net Framework 4.0. If SEEG
software fails to launch properly, please check whether Microsoft .Net
Framework 4.0 or higher versions was installed in the operation system.
If your PC does not install Microsoft .Net Framework 4.0 or higher versions,
please download from Microsoft website. Please click
to watch the
tutorial video (from 2:03).
2.2
Connecting to the ECG
For connecting the ECG device under to the USB module, use the “ECG breakout box”
provided. The ECG breakout box includes:
(1) RA~V6 terminals: Total 10 terminals, corresponding to the 10 electrodes
of 12-lead ECG (or fewer lead channel ECG)
(2) GND terminal: The terminal connected to ground.
(3) CMRR Imbalance with DC: The terminal is only used with CMRR 2.0 for
CMRR testing purpose.
Alternately the ECG device under test can be directly connected to the USB module
using a male D15 connector. The pin outs are:
1-RA
4
–
RL 7
–
V4
10 - V1
2-LA
5
–
V6 8
–
V3
11- NC
3-LL
6
–
V5 9
–
V2
12- GND
2.3
Environment, noise reduction
A noise free environment is necessary for testing ECG equipment. This can be
achieved relatively easily by (a) using a metal bench or metal sheet underneath the
ECG device under test and the WhaleTeq SECG test unit, and (b) connecting SECG
GND terminal to the sheet and also the frame ground (or EP terminal) of the ECG
device under test: