22
Single Channel ECG Test System | User Manual
required by IEC 60601-2-27, 100bpm according to IEC
60601-2-51).
Single/Double
pulses, 150ms and
250ms advanced
This group selects whether single or double pulses are
required according to IEC 60601-2-27. If double pulses are
required, they can be 150 or 250ms advanced.
2.6.7
Output graphic display
The output display provides an image similar to that provided by ECGs. The
sensitivity of the display range may be set at 4mm/mV, 10mm/mV or 20mm/mV to
cover the full range of waveforms offered by the system. The time rate is fixed.
The output display uses the same data as used in the DAC output and serves as a
cross check of the selected waveform, and also allows the user to view the original
waveform as filters in the ECG device under test can substantially alter the waveform.
Pacing pulses are shown in purple.
For Version 4.0, the output range is fixed at ±5mV.