Westlab 663-575 Скачать руководство пользователя страница 9

9

PRODUCT MANUAL

1800 358 101

[email protected]

www.westlab.com.au

+64 9 553 3677

[email protected]

www.westlab.co.nz

646 568 5391

[email protected]

www.westlab.com

1 877 822 1455

[email protected]

www.westlab.com

Australia

New Zealand

United States

Canada

Setting Temperature:

Initially on every start-up, the temperature function will be in OFF state. To set temperature, 

single press ‘TEMP’ to select the temperature value. The selected temperature value is saved 

after the display blinks several times.

Temperature value can be set from ambient temperature to 320°C. Once the temperature 

value is selected, press ‘+/-’ to increase or decrease the temperature value. The temperature 

value is saved after the display blinks several times.

The user can select the unit of temperature (°C) or (°F) by pressing ‘°C/°F’. As per the selected 

unit, the actual value and the set value of temperature will change accordingly.

To operate the device in stirring mode only, select the speed and time as mentioned in 

the previous sections. For temperature, single press the temperature button to select 

the temperature and then press ‘+/-’ until the SET temp display shows ‘Abc’ (Ambient 

Temperature). Finally, press the ‘START’ button to start the stirrer. 

Setting ATS Mode:

ATS Mode is for the auto start of the machine. If ATS is active upon power failure during 

stirring, the machine starts the function with similar parameters once the power resumes.

To activate ATS mode, press the ‘TIME/ATS’ key until the display shows ‘ATS ON’.  

ATS mode will work in either Condition Infinite or Set Timer.

To exit from ATS mode during operation, press ‘TIME/ATS’ key until the display shows  

‘ATS OFF’. If the run is completed for the set timer, the ATS automatically deactivates  

and will need to be reactivated with every new run.

Setting Lock/Unlock Keypad Function:

The keypad can be locked to avoid any accidental or tampering with the setting of the 

parameters. The lock can only be activated during the operation of the device. After pressing 

the ‘START’ button, press the ‘+/-’ buttons together. One can lock the keypad which displays 

as ‘LOC’ on the display panel. This means that only the ‘START/STOP’ button will remain 

functional.

After pressing the ‘STOP’ button in ‘LOC’ condition during operation, the START/STOP button 

will also be locked. To start the operation again, press the ‘Unlock’ button.

In ATS Mode, the lock mode will be unlocked by only two ways. Either by pressing ‘+/-’ buttons 

together or if the unit is restarted after the ATS operation is over. While the ATS operation is 

running the lock mode will be remain ON even if there is a power cut-off & the unit restarts.  

Or else the buttons stay ‘Unlocked’ in all conditions.

Pulse Mode:

The pulse mode functions as Run-Pause-Run. The default pulse mode is always in 

OFF condition.

To activate pulse mode, long press the ‘SPEED/PULSE’ button and the display will show ‘Pulse 

On’. By long pressing the ‘SPEED/PULSE’ button again, it will deactivate the pulse mode and 

the display will show ‘Pulse Off’ even during operation.

Fix time for PULSE mode will be always 30 seconds. By default, the motor runs clockwise. 

Pulse mode can be turned ON/OFF while stirrer is running. Pulse mode will work at any  

SET RPM which is in the range of (200-2200). Time counting of pulse time will counting  

from beginning.

Содержание 663-575

Страница 1: ...1 PRODUCT MANUAL Product Manual Advanced Hot Plate Stirrer Code 663 575 ...

Страница 2: ...ls and universities 1 3 Features Powerful DC Motor Delivers Constant Speed Under Varying Load Conditions Different Setup Plates SS Top with Chrome Plating SS Top with Ceramic Coating Aluminium Top Rapid Heat Up Heats 1L Beaker of Water to 100 C in under 25 Minutes Large Clear Display Showing Critical Parameters for Easy Viewing Microprocessor Controlled Last Run Memory Feature Temperature Setting ...

Страница 3: ...s 1 Stirring Capacity 10L Speed Control 10RPM step Temperature Range Room temperature to 320 C Temperature Control 1 C step Temperature Accuracy K 1 Temperature Safe 345 C Timer Range 1 to 999 minutes infinite mode Permissible Ambient Temperature 5 40 C Permissible Relative Humidity 80 Stirrer Bar Length 25mm recommended Input Voltage 220V 50Hz Heat Output 600W Motor Rating Input Output 5 7W Total...

Страница 4: ...ware of the possible effects of magnetic field on pacemakers data media etc 4 Do not touch the hotplate surface when the hotplate temperature exceeds 50 C Pay attention to the residual heat after switching off the unit 5 Properly lift the device with both hands while moving or installing Also the device should only be moved from its position once it reaches room temperature 6 Make sure that the po...

Страница 5: ...e power supply must be within the specified rating 18 The top surface will be hot after any heating operation Do not touch the top surface until the hot LED indication turns off 19 Process pathogenic materials in a closed vessel under a suitable extractor hood Do not operate the instrument in explosive atmospheres near hazardous substances or under water 20 The chemical reaction of PTFE can occur ...

Страница 6: ...d surface ensure that all the four legs stand on the surface firmly Avoid installing on a slippery or surface prone to vibration Ideal ambient temperature is 25ºC 1ºC avoid placing the unit in direct sunlight Keep clearance of at least 20cm from all side to increase the cooling efficiency Keep the unit away from heated solution to avoid sample temperature issues Do not place the equipment at a pla...

Страница 7: ...ulse Press Speed to select the speed value Then press to set the speed value in RPM Long press Speed Pulse for pulse mode 3 Time ATS Press Time ATS to select the time value Then press to set the time value in minutes Long press Time ATS for ATS mode 4 Temp Selection Press C F to change the unit in which temperature is displayed 5 Increment Press to increase values 6 Decrement Press to decrease val...

Страница 8: ...perature The maximum value of safe temperature can be set up to 345 C After selecting the desired safe temperature value the value gets saved automatically after blinking five times in the display panel Setting Time The time in the display will show in minutes The stirrer timer set for operation as in between 1 minute to 999 minutes and or infinite mode Infinite mode is seen as To set time press t...

Страница 9: ... Infinite or Set Timer To exit from ATS mode during operation press TIME ATS key until the display shows ATS OFF If the run is completed for the set timer the ATS automatically deactivates and will need to be reactivated with every new run Setting Lock Unlock Keypad Function The keypad can be locked to avoid any accidental or tampering with the setting of the parameters The lock can only be activa...

Страница 10: ...side of the instrument The PROBE LED will glow indicating that probe is selected and an ACTUAL display will automatically show the temperature of probe Never remove or insert the PT 1000 probe while the equipment is in operation The user cannot change the plate or probe mode using the key By removing the probe the device will switch to plate mode remove after complete power off To insert or remove...

Страница 11: ...nnection No Heating Remove the probe and plug it back in correctly Er6 Heater is malfunctioning Heater failure No Heating Change the heater assembly WARNING Only to be carried out by authorised service personnel Er7 Ambient temperature sensor failure malfunctioning internal temperature of the unit is more than 80 C No Heating Switch off the device and run it again after 30 minutes If the error con...

Отзывы: