11.  Reduce the speed:

• 

If the medium splashes out of the vessel due to high speed

• 

If the appliance is not running smoothly

• 

If the container moves on the base plate

12.  Do not move the device when it is connected to a power supply or during its operation.
13.  Ensure that the product is used only for specified operations. It should not be used for 

shaking hazardous or reactive solutions.

14.  Give specific attention to the risks associated with:

• 

Flammable materials

• 

Flammable media with low steam pressure

• 

Overfilling of medium

• 

Incorrect vessel size

• 

Unstable vessels

• 

Glass breakage

15.  Beware of the possibility of contamination or unwanted chemical reactions.
16.  The RTD (PT 1000) sensor must always be immersed in the medium by at least 20mm.
17.  The rating of the power supply must be within the specified rating.
18.  The top surface will be hot after any heating operation. Do not touch the top surface  

until the hot LED indication turns off.

19.  Process pathogenic materials in a closed vessel under a suitable extractor hood.  

Do not operate the instrument in explosive atmospheres, near hazardous  

substances or under water.

20.  The chemical reaction of PTFE can occur when in contact with molten or dissolved 

alkaline earth metals, as well as, with fine partied powders of metals of the 2 and 3 

group of the periodical system at temperatures above 300 to 400°C. Only elementary 

fluorine, chlorine trifluoride and alkaline metals attack PTFE. Halogen hydrocarbons 

have a reversible swelling effect. Only glass coated magnetic rods should be used in 

combination with solute alkali metals or alkaline earth metals or at a temperature  

above 250°C.

21.  Do not cover the device, even partially, with elements such as metallic sheets or plates. 

Ensure that the mounting plate remains clean and uncovered.

22.  The socket must be earthed (protective ground contact).
23.  Do not use the device if the ceramic set-up surface is damaged by scratches,  

splinters or corrosion. A damaged set-up surface could break if used.

Содержание 663-575

Страница 1: ...1 PRODUCT MANUAL Product Manual Advanced Hot Plate Stirrer Code 663 575 ...

Страница 2: ...ls and universities 1 3 Features Powerful DC Motor Delivers Constant Speed Under Varying Load Conditions Different Setup Plates SS Top with Chrome Plating SS Top with Ceramic Coating Aluminium Top Rapid Heat Up Heats 1L Beaker of Water to 100 C in under 25 Minutes Large Clear Display Showing Critical Parameters for Easy Viewing Microprocessor Controlled Last Run Memory Feature Temperature Setting ...

Страница 3: ...s 1 Stirring Capacity 10L Speed Control 10RPM step Temperature Range Room temperature to 320 C Temperature Control 1 C step Temperature Accuracy K 1 Temperature Safe 345 C Timer Range 1 to 999 minutes infinite mode Permissible Ambient Temperature 5 40 C Permissible Relative Humidity 80 Stirrer Bar Length 25mm recommended Input Voltage 220V 50Hz Heat Output 600W Motor Rating Input Output 5 7W Total...

Страница 4: ...ware of the possible effects of magnetic field on pacemakers data media etc 4 Do not touch the hotplate surface when the hotplate temperature exceeds 50 C Pay attention to the residual heat after switching off the unit 5 Properly lift the device with both hands while moving or installing Also the device should only be moved from its position once it reaches room temperature 6 Make sure that the po...

Страница 5: ...e power supply must be within the specified rating 18 The top surface will be hot after any heating operation Do not touch the top surface until the hot LED indication turns off 19 Process pathogenic materials in a closed vessel under a suitable extractor hood Do not operate the instrument in explosive atmospheres near hazardous substances or under water 20 The chemical reaction of PTFE can occur ...

Страница 6: ...d surface ensure that all the four legs stand on the surface firmly Avoid installing on a slippery or surface prone to vibration Ideal ambient temperature is 25ºC 1ºC avoid placing the unit in direct sunlight Keep clearance of at least 20cm from all side to increase the cooling efficiency Keep the unit away from heated solution to avoid sample temperature issues Do not place the equipment at a pla...

Страница 7: ...ulse Press Speed to select the speed value Then press to set the speed value in RPM Long press Speed Pulse for pulse mode 3 Time ATS Press Time ATS to select the time value Then press to set the time value in minutes Long press Time ATS for ATS mode 4 Temp Selection Press C F to change the unit in which temperature is displayed 5 Increment Press to increase values 6 Decrement Press to decrease val...

Страница 8: ...perature The maximum value of safe temperature can be set up to 345 C After selecting the desired safe temperature value the value gets saved automatically after blinking five times in the display panel Setting Time The time in the display will show in minutes The stirrer timer set for operation as in between 1 minute to 999 minutes and or infinite mode Infinite mode is seen as To set time press t...

Страница 9: ... Infinite or Set Timer To exit from ATS mode during operation press TIME ATS key until the display shows ATS OFF If the run is completed for the set timer the ATS automatically deactivates and will need to be reactivated with every new run Setting Lock Unlock Keypad Function The keypad can be locked to avoid any accidental or tampering with the setting of the parameters The lock can only be activa...

Страница 10: ...side of the instrument The PROBE LED will glow indicating that probe is selected and an ACTUAL display will automatically show the temperature of probe Never remove or insert the PT 1000 probe while the equipment is in operation The user cannot change the plate or probe mode using the key By removing the probe the device will switch to plate mode remove after complete power off To insert or remove...

Страница 11: ...nnection No Heating Remove the probe and plug it back in correctly Er6 Heater is malfunctioning Heater failure No Heating Change the heater assembly WARNING Only to be carried out by authorised service personnel Er7 Ambient temperature sensor failure malfunctioning internal temperature of the unit is more than 80 C No Heating Switch off the device and run it again after 30 minutes If the error con...