1-14
Electromagnetic immunity
The Welch Allyn KleenSpec® Corded Illumination System is intended for use in
the electromagnetic environment specified below. The customer or the user of
the Welch Allyn KleenSpec® Corded Illumination System should assure that it
is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Portable and mobile RF
communications
equipment should be used
no closer to any part of the
KleenSpec® Corded
Illumination System,
including cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
d
=(1.17)
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz
to 1 GHz
3 V/m
d
=(1.17)
80 to 800
MHz
P
P
Содержание KleenSpec 789 Series
Страница 1: ...Welch Allyn KleenSpec 78900 and 78950 Corded Illumination System Directions for use...
Страница 4: ......
Страница 22: ......
Страница 40: ......
Страница 58: ......
Страница 76: ......
Страница 94: ......
Страница 98: ...6 4 Is tning af belysningssystemet Model 580 serien Model 590 serien R r...
Страница 112: ......
Страница 130: ......
Страница 148: ......
Страница 152: ...9 4 Valaisimen asentaminen 580 sarjan malli 590 sarjan malli Suojus...
Страница 166: ......
Страница 184: ......
Страница 202: ......
Страница 219: ......