1-13
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
) for 0.5
cycle
40%
U
T
(60% dip in
U
T
) for 5
cycles
70%
U
T
(>95% dip in
U
T
) for 25
cycles
<5%
U
T
(>95% dip in
U
T
) for 5
seconds
<5%
U
T
(>95% dip in
U
T
) for 0.5
cycle
40%
U
T
(60% dip in
U
T
) for 5
cycles
70%
U
T
(>95% dip in
U
T
) for 25
cycles
<5%
U
T
(>95% dip in
U
T
) for 5
seconds
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the KleenSpec® Corded
Illumination System
requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered from an
uninterruptible power
supply or a battery.
Power
frequency (50/
60Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note:
U
T
is the AC mains voltage prior to application of the test level.
Содержание KleenSpec 789 Series
Страница 1: ...Welch Allyn KleenSpec 78900 and 78950 Corded Illumination System Directions for use...
Страница 4: ......
Страница 22: ......
Страница 40: ......
Страница 58: ......
Страница 76: ......
Страница 94: ......
Страница 98: ...6 4 Is tning af belysningssystemet Model 580 serien Model 590 serien R r...
Страница 112: ......
Страница 130: ......
Страница 148: ......
Страница 152: ...9 4 Valaisimen asentaminen 580 sarjan malli 590 sarjan malli Suojus...
Страница 166: ......
Страница 184: ......
Страница 202: ......
Страница 219: ......