1-7
4. Rough dry the illuminator.
5. The illuminator is now ready for disinfection.
Disinfecting
Refer to the Cleaning Table to identify the appropriate disinfection method/prod-
uct for the illuminator and power supply.
Illuminator
1.
Clean or disinfect as described in the table above.
2. Follow the disinfecting wipe manufacturer’s disinfection instructions.
3. Thoroughly rinse illuminator with 22° C to 43° C tap water to remove
disinfectant.
4. Allow to dry.
Power supply
DO NOT IMMERSE THE POWER SUPPLY.
1.
Clean or disinfect as described in the table above.
2. Follow the disinfecting wipe manufacturer’s disinfection instructions on the
cord only.
3. Thoroughly rinse the
cord only
with 22° C to 43° C tap water to remove
disinfectant. Do not immerse the power supply.
4. Allow to dry.
Disposal
There are no harmful materials associated with the KleenSpec Corded
Illumination System.
Device is considered a biohazard after use. Follow good hospital practices and
local ordinances in the disposal of the illuminator.
There are no risks associated with the disposal of this device after its useful life if
it is cleaned and/or disinfected per the instructions in this manual.
Содержание KleenSpec 789 Series
Страница 1: ...Welch Allyn KleenSpec 78900 and 78950 Corded Illumination System Directions for use...
Страница 4: ......
Страница 22: ......
Страница 40: ......
Страница 58: ......
Страница 76: ......
Страница 94: ......
Страница 98: ...6 4 Is tning af belysningssystemet Model 580 serien Model 590 serien R r...
Страница 112: ......
Страница 130: ......
Страница 148: ......
Страница 152: ...9 4 Valaisimen asentaminen 580 sarjan malli 590 sarjan malli Suojus...
Страница 166: ......
Страница 184: ......
Страница 202: ......
Страница 219: ......