1-6
Cleaning and Disinfecting
Cleaning
Wipe
KleenSpec Corded Illumination System
Refer to Cleaning Table to identify the appropriate cleaning product.
1.
Disconnect the power supply from the wall outlet.
2. Follow the disinfecting wipe manufacturer’s instruction and wipe cord or
illuminator completely to remove any residue.
3. This process is a rough clean only. Refer to disinfection for further
processing.
Soak
Refer to Cleaning Table to identify the appropriate cleaning product.
KleenSpec Corded Illuminator
1.
Disconnect the power supply from the wall outlet.
2. Follow the cleaning solution manufacturer’s instructions and soak the
illuminator as instructed. Solution temperature is 22° C to 43° C. Make sure
to keep the power supply out of the soak solution. Do not immerse the
power supply.
3. Scrub any area of visible debris with a soft brush and thoroughly rinse the
illuminator with 22° C to 43° C tap water to remove the cleaning solution. Do
not rinse the power supply; wipe only.
KleenSpec Corded
Illuminator
Power supply
Cleaning
CaviWipe
(Wipe)
Yes
Yes
Enzol (Soak)
Yes
No*
Metrizyme
(Soak)
Yes
No*
Disinfecting
Caviwipe
Intermediate Level
Intermediate
Level
* Do not immerse the power supply plug. Follow instructions per this
manual.
CAUTION: Lamp must remain installed during all cleaning and
disinfecting processes.
CAUTION: Lamp must remain installed during all cleaning and
disinfecting processes.
Содержание KleenSpec 789 Series
Страница 1: ...Welch Allyn KleenSpec 78900 and 78950 Corded Illumination System Directions for use...
Страница 4: ......
Страница 22: ......
Страница 40: ......
Страница 58: ......
Страница 76: ......
Страница 94: ......
Страница 98: ...6 4 Is tning af belysningssystemet Model 580 serien Model 590 serien R r...
Страница 112: ......
Страница 130: ......
Страница 148: ......
Страница 152: ...9 4 Valaisimen asentaminen 580 sarjan malli 590 sarjan malli Suojus...
Страница 166: ......
Страница 184: ......
Страница 202: ......
Страница 219: ......