1-12
Electromagnetic immunity
The Welch Allyn KleenSpec® Corded Illumination System is intended for use in
the electromagnetic environment specified below. The customer or the user of
the Welch Allyn KleenSpec® Corded Illumination System should assure that it
is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV
differential
mode
± 2kV
common
mode
±1 kV
differential
mode
± 2kV
common
mode
Mains power quality should
be that of a typical
commercial or hospital
environment.
Содержание KleenSpec 789 Series
Страница 1: ...Welch Allyn KleenSpec 78900 and 78950 Corded Illumination System Directions for use...
Страница 4: ......
Страница 22: ......
Страница 40: ......
Страница 58: ......
Страница 76: ......
Страница 94: ......
Страница 98: ...6 4 Is tning af belysningssystemet Model 580 serien Model 590 serien R r...
Страница 112: ......
Страница 130: ......
Страница 148: ......
Страница 152: ...9 4 Valaisimen asentaminen 580 sarjan malli 590 sarjan malli Suojus...
Страница 166: ......
Страница 184: ......
Страница 202: ......
Страница 219: ......