Service Manual
CP 100 and CP 200 Electrocardiograph
67
Replacing the LCD Assembly (CP 100)
Procedure
1.
Open unit using the
directions.
2.
Disconnect the flat cable from the
keyboard. See
3.
Remove the 4 screws securing the LCD
bracket. See
.
4.
Gently lift out the LCD bracket.
5.
Gently lift out the LCD assemby.
Figure 57. CP 100 Top Cover
Screws
LCD Assembly
LCD Bracket
Flat Cable