56
Disassembly and Repair Procedures
Welch Allyn CP 100 and CP 200 Electrocardiograph
5.
Disconnect the paper detector harness
from the back of the printhead assembly.
See
6.
Disconnect the ground connection using a
¼” nut driver. See
7.
Remove the 2 shoulder screws from the
opposite end of the printhead using a 3/16”
nut driver.
8.
Remove the 2 screws holding the Que
sensor Assembly. See
9.
Carefully remove the printhead connectors
from both ends of the printhead.
Figure 43. Disconnecting Paper Detector Harness
and Ground Connections
Note
The printhead does not come with the Antistatic brushes. Order the Antistatic
brushes separately. See
Procedure
Ground Connection
Paper Detector Harness
Screws holding the Que
Sensor Assembly
Содержание CP 100
Страница 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Страница 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Страница 53: ...5 47 Disassembly and Repair Procedures Introduction 48 Screws 48 Torque Specs 48 Disassembly Procedures 49 ...
Страница 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Страница 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Страница 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 95: ......
Страница 96: ...Material Number 704340 Rev D PO Box 220 Skaneateles Falls NY 13153 0220 USA 800 535 6663 www welchallyn com ...