30
Functional Verification
Welch Allyn CP 100 and CP 200 Electrocardiograph
Functional Verification Overview
This section describes the procedure for a complete functional test to support
recommended preventive-maintenance schedules.
A checklist of the functional tests is provided on
. It is recommended that you
print a copy of the checklist each time you perform the functional verification procedure,
so you can record and save the test results. If the electrocardiograph ever requires
service, the records of test results often facilitate troubleshooting.
Functional verification does not require opening the device case.
Full Functional Check
Whenever the electrocardiograph is serviced or problems are suspected, Welch Allyn
recommends a Functional Check.
Equipment Required
Operating/Non-Printing Mode Current Test
When the unit is powered on, the AC is disconnected and the LCD on, the unit’s current
is:
1.
Unplug unit from line power
2.
Connect a DC power supply, adjusted for 6.4V DC ± .05, in place of the battery.
3.
Verify the current draw is within limits as indicated for unit after powering the unit up.
WARNING
Only qualified service personnel should perform a full functional
checkout procedure.
Commercially Available General-purpose/Medical Test Equipment
Item
Manufacturer
Power Supply
Variable, 0 – 8 VDC, 5.0A
(minimum), with voltage and current indicators
AC withstand voltage (hi-pot) tester Associated Research 3605 or equivalent
Electrical Safety Analyzer w/ECG
Fluke DNI 232D or equivalent
Computer
PIII or better, CardioPerfect installed, available USB port
SD Memory Card
64Mb or Greater
USB Cable
A – A type (WA P/N 401435)
Spirometry Dongle
Serial Version (W/A P/N 703552)
CP 100 Current
CP 200 Current
125 to 250 mA @ 6.4V DC ± .05
675 to 825 mA @ 6.4V DC ± .05
Содержание CP 100
Страница 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Страница 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Страница 53: ...5 47 Disassembly and Repair Procedures Introduction 48 Screws 48 Torque Specs 48 Disassembly Procedures 49 ...
Страница 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Страница 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Страница 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 95: ......
Страница 96: ...Material Number 704340 Rev D PO Box 220 Skaneateles Falls NY 13153 0220 USA 800 535 6663 www welchallyn com ...