36
Functional Verification
Welch Allyn CP 100 and CP 200 Electrocardiograph
Table 2. Print Verification Check List
Figure 29. Test Printout Example
Point
Description
Point A
Verify the distance of the arc on the circle from the top edge of the paper is 37mm +/-
3mm
Point B
Verify the distance of the arc on the circle from the left edge of the paper is 45mm +/-
5mm
Point C
Verify line is straight and without breaks from top to bottom
Point D
Verify thick black line is Uniform in Blackness without breaks from top to bottom
Point E
Verify line is straight and measures 200mm +/- 4mm
Point F
Verify the print is readable
Point G
Verify the lead in and output trail of the cue sensor are between 19 and 82mm down
from the 200mm line measured in point E
Point H
Verify the cue sense pattern is between 97 and 135mm down from the 200mm line
measured in point E
Содержание CP 100
Страница 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Страница 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Страница 53: ...5 47 Disassembly and Repair Procedures Introduction 48 Screws 48 Torque Specs 48 Disassembly Procedures 49 ...
Страница 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Страница 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Страница 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 95: ......
Страница 96: ...Material Number 704340 Rev D PO Box 220 Skaneateles Falls NY 13153 0220 USA 800 535 6663 www welchallyn com ...