Disassembly and Repair Procedures
Welch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Battery or Fuse
Procedure
1.
Turn the electrocardiograph up-side-down.
2.
Unscrew and remove the battery door. See
Figure 46. Removing the Battery Door
3.
Lift out the battery. See
.
Figure 47. Lifting Out the Battery
4.
Inspect the fuse. See
•
If the “Z” wire is intact, go to
.
•
If the “Z” wire is broken or dark,
replace the fuse. See
Figure 48. Battery Fuse Plugged In
Battery door
Battery
Fuse
Содержание CP 100
Страница 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Страница 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Страница 53: ...5 47 Disassembly and Repair Procedures Introduction 48 Screws 48 Torque Specs 48 Disassembly Procedures 49 ...
Страница 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Страница 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Страница 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 95: ......
Страница 96: ...Material Number 704340 Rev D PO Box 220 Skaneateles Falls NY 13153 0220 USA 800 535 6663 www welchallyn com ...