Service Manual
CP 100 and CP 200 Electrocardiograph
49
Disassembly Procedures
Opening the Unit
WARNING
Disconnect ALL power from the unit before repairing or
disassembling unit. Unplug the power cord and disconnect the battery at the
fuse in the battery compartment. See
Replacing the Battery or Fuse (page 58)
Procedure
Figure 31. Removing the Screws
1.
Place unit on a clean, static mat.
2.
Turn the unit up-side-down
3.
Remove the 8 screws around the outside
rim of the unit. See
.
4.
Using both hands, hold the upper case and
lower case of the unit together, turn the
unit upright.
5.
Slowly lift the upper case, and disconnect
the ribbon cable from the upper case.
6.
Lift off the top cover. See
Figure 32. Disconnecting the Ribbon Cable
Screws
Ribbon Cable
Upper Case
Lower Case
Содержание CP 100
Страница 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Страница 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Страница 53: ...5 47 Disassembly and Repair Procedures Introduction 48 Screws 48 Torque Specs 48 Disassembly Procedures 49 ...
Страница 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Страница 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Страница 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Страница 95: ......
Страница 96: ...Material Number 704340 Rev D PO Box 220 Skaneateles Falls NY 13153 0220 USA 800 535 6663 www welchallyn com ...