5. The monitor sounds a tone when the final temperature is reached (in approximately 10 to 13
seconds). The temperature frame continues to display the temperature in degrees Fahrenheit
and degrees Celsius even after the probe is returned to the probe well.
Note
To switch to the Direct mode, touch
after the Predictive
mode measurement is acquired. The temperature frame (in the
lower-left corner) changes to "MODE: Direct..." as it switches to the
Direct mode. The monitor sounds a tone to indicate the start of a
Direct measurement. Once you are in the Direct mode, continue to
separate the buttocks and hold the probe in place throughout the
measurement process.
Note
The monitor does not retain Direct mode temperatures in memory.
Therefore, it is important to note the temperature before removing
the probe from the measurement site and then manually record it in
the patient record.
6. Remove the probe after the temperature measurement is complete and firmly press the eject
button on the top of the probe to release the probe cover.
7. Return the probe to the probe well.
8. Wash your hands to reduce the risk of cross-contamination.
Braun ThermoScan
®
PRO 4000 thermometer and accessory dock
The thermometer and accessory dock enable you to transfer an ear temperature measurement to
the monitor. The dock also charges the thermometer battery.
Read the thermometer manufacturer's directions for use before attempting to configure, use,
troubleshoot, or maintain the thermometer.
WARNING
Liquids can damage electronics inside the thermometer. Prevent
liquids from spilling on the thermometer. If liquids are spilled on the thermometer,
dry off the thermometer with a clean cloth. Check for proper operation and
accuracy. If liquids possibly entered the thermometer, remove the thermometer
from use until it has been properly dried, inspected, and tested by qualified service
personnel.
Caution
Probe covers are disposable, nonsterilized, and single-use. The
thermometer is also nonsterilized. Do not autoclave the thermometer and probe
covers. Ensure that probe covers are disposed of according to facility requirements
or local regulations.
Caution
The thermometer has no user-serviceable parts. If service is required,
call your nearest Welch Allyn Customer Service or Technical Support facility.
76
Patient monitoring
Welch Allyn Vital Signs Monitor 6000 Series™
Содержание Connex Vital Signs Monitor 6000 Series
Страница 1: ...Welch Allyn Vital Signs Monitor 6000 Series Directions for use ...
Страница 6: ...vi Contents Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 8: ...2 Introduction Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 14: ...8 Screen elements Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 22: ...16 Controls indicators and connectors Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 30: ...24 Setup Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 40: ...34 Navigation Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 44: ...38 Profiles Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 64: ...58 Alarms Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 92: ...86 Patient monitoring Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 98: ...92 Maintenance and service Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 106: ...100 Specifications Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 134: ...128 Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 143: ......