Button name
Button image
Description
Direct mode
Touch the button to enter Direct mode.
Configure temperature alarms
Follow these steps to set alarm limits for temperature measurement.
1. Verify that you are using the Monitor profile, which contains the Alarms tab.
2. Touch the
Alarms
tab.
3. Touch the
Temperature
tab.
4. Enter the desired upper and lower alarm limits for temperature using the up/down arrow keys
or the keypad.
5. Touch the
Home
tab.
The new alarm settings display in the Alarm Limit control button.
SureTemp
®
Plus temperature module
The temperature module uses a thermistor thermometer design and a predictive algorithm to
calculate patient temperatures in the Predictive mode.
WARNING
Patient injury risk. Prior to taking a temperature, instruct the patient
not to bite down on the probe as patient injury and damage to the probe may result.
WARNING
Patient injury risk. Do not exceed the recommended temperature
measurement durations in Direct mode. Continuous measurement durations of 3
minutes at the oral and rectal sites and 5 minutes at the axillary site are
recommended for accurate measurement. Do not continuously measure beyond 10
minutes in any mode.
WARNING
Probe covers are single-use only. Re-use of a probe cover may result
in spread of bacteria and cross-contamination.
WARNING
Patient injury risk. Use only Welch Allyn single-use disposable probe
covers. Never take a temperature measurement without a single-use probe cover
securely attached. Failure to use a probe cover can cause patient discomfort from a
heated probe, patient cross-contamination, and inaccurate temperature readings.
WARNING
Patient illness may result from improper use of oral and rectal
temperature probes. Using the incorrect probe may also produce inaccurate
measurements.
•
Use only oral probes, identified by a blue ejection button at the top of the probe, to
take oral and axillary temperatures.
•
Use only rectal probes, identified by a red ejection button at the top of the probe, to
take rectal temperatures.
70
Patient monitoring
Welch Allyn Vital Signs Monitor 6000 Series™
Содержание Connex Vital Signs Monitor 6000 Series
Страница 1: ...Welch Allyn Vital Signs Monitor 6000 Series Directions for use ...
Страница 6: ...vi Contents Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 8: ...2 Introduction Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 14: ...8 Screen elements Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 22: ...16 Controls indicators and connectors Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 30: ...24 Setup Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 40: ...34 Navigation Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 44: ...38 Profiles Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 64: ...58 Alarms Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 92: ...86 Patient monitoring Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 98: ...92 Maintenance and service Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 106: ...100 Specifications Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 134: ...128 Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 143: ......