Troubleshooting
This section presents tables of technical alarm and information messages to help you troubleshoot
issues on the monitor.
When the monitor detects certain events, a message appears in the Device Status area at the top
of the screen. Message types include the following:
•
Information messages appear on a blue background.
•
Low- and medium-priority alarms appear on an amber background.
•
High-priority alarms appear on a red background.
Technical alarm messages are low priority unless noted in the Message column.
You can dismiss a message by touching the message on the screen, or, for some messages, you
can wait for the message to time out.
To use these tables, locate the message that displays on the monitor in the left column of the table.
The remainder of the row explains possible causes and suggests actions that can resolve the
issue.
Note
Instructions to "Call for service" in the following tables mean that you should contact
qualified service personnel in your facility to investigate the issue.
NIBP messages
Message
Possible cause
Suggested action
NIBP air leak; check cuff and
tubing connections.
The NIBP module has an air leak
Check the cuff and tubing
connections.
Clear the alarm and retry
NIBP.
NIBP not functional. Call for
service.
The ambient temperature is out of range
Use the monitor in the
specified temperature range.
Call for service.
Unable to determine NIBP;
check connections; limit patient
movement.
The NIBP module experienced a motion
artifact
Check connections; limit
patient movement.
Clear the alarm and retry
NIBP.
Unable to determine NIBP;
check connections and tubing.
The NIBP tubing has a kink
Check the connections and
tubing for kinks.
121
Содержание Connex Vital Signs Monitor 6000 Series
Страница 1: ...Welch Allyn Vital Signs Monitor 6000 Series Directions for use ...
Страница 6: ...vi Contents Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 8: ...2 Introduction Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 14: ...8 Screen elements Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 22: ...16 Controls indicators and connectors Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 30: ...24 Setup Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 40: ...34 Navigation Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 44: ...38 Profiles Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 64: ...58 Alarms Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 92: ...86 Patient monitoring Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 98: ...92 Maintenance and service Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 106: ...100 Specifications Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 134: ...128 Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 143: ......