Electrical interference immunity
55
Guidelines and manufacturer’s declaration – electromag-
netic interference immunity
This device is intended for use in electromagnetic envi-
ronments as described below. The customer or operator is
responsible for observing that the device is used in such an
environment.
Interference
immunity
tests
Test level
in accor-
dance with
IEC 60601
Degree
of con-
formity
Electromagnetic
environment
Electrostatic
discharge
(ESD) in accor-
dance with IEC
61000-4-2
± 8 kV
contact
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV air
Confor-
mant
Floor surfaces should
be wood, concrete
or ceramic tile. If the
floor has a synthetic
covering, the relative
humidity must be at
least 30 %.
Содержание 96373
Страница 3: ...A 3 1 2 3 4 5 ...
Страница 4: ...B C 4 14 13 12 11 6 7 8 9 10 15 16 ...
Страница 5: ...D 5 E F 2 28 28 27 26 25 24 17 18 19 20 21 23 22 ...
Страница 46: ...Adhesive electrodes 46 Example for application of adhesive electrodes 1 57 1 57 in 40 40 mm ...
Страница 109: ...Electrodos adhesivos 109 Ejemplos de uso de electrodos adhesivos de 1 57 1 57 in 40 40 mm ...