Repairguide for VENTImotion/VENTIlogic
35
14. Insert the rechargeable battery connecting
cable into the groove on the power board.
Caution!
Be absolutely sure that the power board/mains
supply unit connecting cable 45 is between the
head of the screw and the box as well as in the
groove of the power board. Ensure that the re-
chargeable battery cable is in the groove. Other-
wise the cable may be trapped and damaged
when the device is closed.
15. Close the device (see "7.3 Close the device"
16. Test the device (see "4. Test the device" on
7.6 Replace box and filter holder
Tools and equipment required
• ESD protected work-station
• 3 mm Allen key
1. Open the device (see "7.2 Open device" on
2. Remove the power board (see "7.5 Replace
3. Remove the mains supply unit (see "7.4 Re-
place mains supply unit" on page 33).
4. Release the pressure-side hose
(connection
to device outlet) from the box.
5. Lift box
part of the housing together.
6. Disconnect the box and the filter holder from
one another by undoing the intake-side hose