40
Repairguide for VENTImotion/VENTIlogic
8. Remove spring
from the control board and
on control board" on page 42). You need the
metal spring for the new control board.
If it is unnecessary to replace the display, con-
tinue from step
.
10. Insert a new display
.
Caution!
The display may not tilt and must be straight in
the guides.
11. Insert a new battery
er of the new control board and screw on
spring
(see "7.9 Replace battery on control
12. Put the control board in the top part of the
housing and attach it with the four screws
13. Reattach the pressure measurement hoses
to the pressure sensors.
Ensure that you do not switch the hoses:
connect
the connection of the flow sensor marked
Hi
to
the hose connector marked
Hi
at the device
outlet.
14. Attach the flow measurement hoses
to the
flow sensor.
Ensure that you do not switch the hoses:
connect
the connection of the flow sensor marked
Lo
to
the hose connector marked
Lo
at the device
outlet.
15. To prevent hoses kinking, position flow mea-
Display buffer
Lo
Hi