38
Repairguide for VENTImotion/VENTIlogic
12. Insert a new lid cushioning insert
cushioning insert
in
box
Take care with the assembly position.
– Insert the labyrinths as shown.
– Insert motor frame
so that the cutout for
the fan cable is flush with the cutout in the
box. Otherwise the fan cable might be
damaged when the box is closed.
13.
Tip!
Put motor frame
with box cushioning insert
, in box
Clean the fan and then disinfect it using
MIKROZID LIQUID (follow the manufacturer's
instructions).
14. Fit the fan flap with two new screws
the maintenance kit.
15. Insert fan
16. Apply a little 70 % isopropanol to the circum-
ference of decoupling tube
and fit it in the
appropriate bore of motor frame
The groove of the hose must engage right
round the circumference of the bore.
17. Put the lid on box
Ensure that the lid is fitted in the correct position.
The cable harness of the fan must run through the
cutout in the lid and may not be trapped! The
cable must protrude between 265 and 270 mm.
18. Refit the box and filter holder (see "7.6 Re-
place box and filter holder" on page 35)
19. Refit the power board (see "7.5 Replace
20. Refit the mains supply unit (see "7.4 Replace
mains supply unit" on page 33).
21. Close the device (see "7.3 Close the device"
22. Test the device (see "4. Test the device" on
Содержание BiLevel-ST VENTImotion WM 24800