Repairguide for VENTImotion/VENTIlogic
43
4. Remove box and filter holder (see "7.6 Re-
place box and filter holder" on page 35).
5. Remove the control board (see "7.8 Replace
control board/display" on page 39).
6. Pull the pressure-side hose off device outlet
.
Dispose of the pressure-side hose (only when
servicing or changing patients).
7. Pull flow measurement hoses off the flow sen-
sor. Dispose of flow measurement hoses
8. Pull pressure measurement hoses
off the de-
vice outlet. Dispose of pressure measurement
hoses.
9. Remove device outlet
the housing
. To do so, lift the tab slightly us-
ing a screwdriver and take out the device out-
let.
10. Remove any adhesive residues from the hous-
ing.
11. Place a spot of glue on the lock:
Ø
≤
1 mm; glue WM 14946 (Loctite 4601)
or WM 14952 (Loctite 4031).
12. Fit a new device outlet
housing
13. Put new pressure measurement hoses
relevant connectors of device outlet
14. Put new flow measurement hoses
on the rel-
evant connectors of device outlet
15. Refit the control board (see section 7.8 on
16. Refit the box and filter holder (see section 7.6
17. Refit the power board (see section 7.5 on
18. Refit the mains supply unit (see section 7.4 on
19. Close the device (see section 7.3 on
20. Test the device (see "4. Test the device" on
glued at
certain spots