Repairguide for VENTImotion/VENTIlogic
47
7.14 Replace top part of housing
Tools and equipment required
• ESD protected work-station
• 3 mm Allen key
• Phillips screwdriver, size 1
1. Open the device (see section 7.2 on
2. Remove the mains supply unit (see section 7.4
3. Remove the power board (see section 7.5 on
4. Remove box and filter holder (see section 7.6
5. Remove the control board (see section 7.8 on
6. Remove the device outlet (see section 7.10 on
7. Remove the lock washers from pins
8. Push pins
out of the hinges and take handle
off the top part of the housing.
9. Remove mains input connector
You have now dismantled all parts. Now start as-
sembling into the new top part of the housing
10. Take the new top part of the housing
hand and insert handle
in the top part of the
housing.
11. Push pins
into the hinges and fit the lock
washers.
12. Fit the device outlet (see section 7.10 on
13. Refit the display and the control board (see
14. Refit the box and filter holder (see section 7.6
15. Refit the power board (see section 7.5 on
16. Refit mains input connector
17. Refit the mains supply unit (see section 7.4 on
18. Close the device (see section 7.3 on
19. Test the device (see "4. Test the device" on
Obe