Wealtec CB-1 Скачать руководство пользователя страница 16

Block Cooler V1.0 

 14

F. Troubleshooting 

 

Note:

 If you notice any malfunction in your block cooler system or it operates in an unusual 

way you may initially attempt to solve the problem by following the general symptoms or 

error messages, this is when the system encounters a known problem and the corrective 

action or solutions given in the following tables below in this section.   

 

Any other than the given solution such as servicing attempts will void the warranty, void the 

regulatory certifications and create potential safety hazard. Please contact your local 

WEALTEC representative if you are unable to solve the problem. 

 

 

F-1. General symptom 
 

Symptom 

Action and possible solution 

Fan is not spinning 

LED & Display do not light up 

Front panel power on indicator off 

Check main AC power source 

Replace a new same configuration fuse and try again. 

Check if power switch malfunction. 

Check if the system displays any errors. 

Contact your local Wealtec representative. 

Fan is not spinning 

LED & Display do not light up 

Front panel power on indicator off 

Check the main AC power source. 

Check if the system displays any errors. 

Contact your local Wealtec representative. 

Fan is not spinning 

LED & Display do not light up 

Front panel power on indicator on 

Contact your local Wealtec representative. 

Intermittent fan, LED or display 

Contact your local Wealtec representative. 

Requires higher offset value 

(More or less than ±12.5

℃)

 

Contact your local Wealtec representative. 

Fuse blown 

Change a new same configure fuse and try again. 

Contact your local Wealtec representative. 

 

 

Содержание CB-1

Страница 1: ...Block Cooler System CB 1 Installation and Operation Manual Version 1 0 Item 03060 This instrument is intended for laboratory use only...

Страница 2: ...pecifications B 2 Product Description B 2 1 Block Cooler System Hardware Overview C Installation of Block Cooler System CB 1 9 C 1 Package List C 2 Installation of Block Cooler System C 2 1 Installati...

Страница 3: ...rp A 1 Warranty Block Cooler system is warranted to be free from defects in materials or workmanship for a period of one year from the original invoice date under normal usage Any defects occurring du...

Страница 4: ...tion Do not apply more or lower than the rated voltage specified on the instrument or it may cause damages to the instrument Do not splash water or liquid onto the block cooler system the spilled wate...

Страница 5: ...ded Applications 0 C Replace the ice buckets for samples enzymes storage 4 C Sample cooling storage during operation i e Store nucleic acid at 4 C in long PCR at final stage DNA purification kit rinsi...

Страница 6: ...etry of Apoptosis Incubate the plug at 37 C in Pulsed Field Gel Electrophoresis For gel digestion pre treatment before MALDI TOF MS analysis 42 C The reverse transcription temperature by RT PCR Enzyme...

Страница 7: ...rose dissolving Denatured sample i e Pre heating protein molecules before SDS PAGE running Customize Temperature Between 0 75 C Specific incubation temperatures for chemistry molecular biology experim...

Страница 8: ...Block capacity 1 Operating conditions Temperature 0 26 Humidity 10 to 90 R H Non condensing Aluminum block Block type 48 Well aluminum block Well type 8 x 0 2 ml 21 x 0 5 ml 15 x 1 5 ml 4 x 2 0 ml Nu...

Страница 9: ...Press this button and adjust the offset value in order to alter the temperature of the block cooler system to the desired temperature as per to users Process indicator Power switch Protection lid Pro...

Страница 10: ...ressing to the desired temperature Up and down key is used to adjust the desired temperature value and the desired mode of the temperature LCD display Displays the temperature value and the mode setti...

Страница 11: ...Block Cooler V1 0 9 C Installation of Block Cooler System CB 1 C 1 Package List Item Quantity Block cooler system CB 1 1 48 Well aluminum block 1 Block protection rubber 1 Block extractor 1...

Страница 12: ...with the minimum size of 200mm L x 304mm W x 184mm H 7 87 x 11 97 x 7 24 2 Place the 48 well aluminum block into the block cooler system and place the block protection rubber on top of the block 3 Th...

Страница 13: ...setting the temperature parameter press Run Stop key once to start running the block cooler Note In general the block cooler system will require several minutes to achieve the desired temperature on...

Страница 14: ...y thermometer is not calibrated accordingly to international standard thermometer Some temperature variation may occur with the usage of the different non calibrated thermometers Caution Maximum adjus...

Страница 15: ...e system achieves the desired temperature 4 and stabilized Ready indicator turned on h When the ready indicator turned on ensure the temperature reading on the users calibrated thermometer is as per t...

Страница 16: ...tion Fan is not spinning LED Display do not light up Front panel power on indicator off Check main AC power source Replace a new same configuration fuse and try again Check if power switch malfunction...

Страница 17: ...ntative Overheat Error The block cooler system is overheated during a run temperature of the system exceeded more than 100 C Ensure the block cooler system operating condition temperature 0 26 Contact...

Страница 18: ...lean water Organic solvents or strong detergents may damage the instrument and should not be used 48 Well aluminum block should be protected from organic solvents and detergents and it should be clean...

Страница 19: ...ption 1093001 CB 1 Block Cooler System with block extractor 48 well aluminum block and block protection rubber 100 240VAC 50 60Hz Optional Accessories Item Description 1093101 48 Well aluminum block f...

Страница 20: ...03060 Wealtec Corp 1885 Meadowvale Way Sparks NV 89431 Tel 1 775 351 2066 Fax 1 775 351 2077 http www wealtec com E mail support wealtec com...

Отзывы: