Block Cooler V1.0 

 13

e.  When the ready indicator turned on, stop the block cooler system.   

 

f.  Press the “Offset” key and by using temperature up and down key adjust the desired 

offset value (The difference between block cooler system temperature and the user’s 

calibrated thermometer). Upon setting the offset value press the “Offset” key again 

 

 Note:

 The maximum adjustable offset range is ±12.5

℃. 

While the system is in the offset function user will not be able to start the system. User 

will need to press the “Offset” key again to quit the offset function and back to main 

menu (Block Cooler CB-1 Set Temp.). 

 

g.  Start the block cooler system. Wait (~15 minutes) until the system achieves the 

desired temperature (4

) and stabilized (Ready indicator turned on). 

 

h.  When the ready indicator turned on ensure the temperature reading on the users 

calibrated thermometer is as per to the previously entered offset value or else repeat 

step (e) – (h). 

 

 Example:

 Assume upon stabilizing the block cooler system temperature, user’s 

calibrated thermometer temperature displays 6.5

  and the block cooler LCD displays 

4

. The offset value that user needed to enter is -2.5

. Operator shall stop the 

system and press offset key and press temperature down key to -2.5

 (6.5

-2.5

4

) then press the “Offset” key again to quit the offset mode. Then start the block 

cooler system and wait until the system temperature stabilize and ensure the reading 

temperature value on the user’s calibrated thermometer is as per to the offset 

adjustment value previously made.   

 

 

 

 

 

 

 

Содержание CB-1

Страница 1: ...Block Cooler System CB 1 Installation and Operation Manual Version 1 0 Item 03060 This instrument is intended for laboratory use only...

Страница 2: ...pecifications B 2 Product Description B 2 1 Block Cooler System Hardware Overview C Installation of Block Cooler System CB 1 9 C 1 Package List C 2 Installation of Block Cooler System C 2 1 Installati...

Страница 3: ...rp A 1 Warranty Block Cooler system is warranted to be free from defects in materials or workmanship for a period of one year from the original invoice date under normal usage Any defects occurring du...

Страница 4: ...tion Do not apply more or lower than the rated voltage specified on the instrument or it may cause damages to the instrument Do not splash water or liquid onto the block cooler system the spilled wate...

Страница 5: ...ded Applications 0 C Replace the ice buckets for samples enzymes storage 4 C Sample cooling storage during operation i e Store nucleic acid at 4 C in long PCR at final stage DNA purification kit rinsi...

Страница 6: ...etry of Apoptosis Incubate the plug at 37 C in Pulsed Field Gel Electrophoresis For gel digestion pre treatment before MALDI TOF MS analysis 42 C The reverse transcription temperature by RT PCR Enzyme...

Страница 7: ...rose dissolving Denatured sample i e Pre heating protein molecules before SDS PAGE running Customize Temperature Between 0 75 C Specific incubation temperatures for chemistry molecular biology experim...

Страница 8: ...Block capacity 1 Operating conditions Temperature 0 26 Humidity 10 to 90 R H Non condensing Aluminum block Block type 48 Well aluminum block Well type 8 x 0 2 ml 21 x 0 5 ml 15 x 1 5 ml 4 x 2 0 ml Nu...

Страница 9: ...Press this button and adjust the offset value in order to alter the temperature of the block cooler system to the desired temperature as per to users Process indicator Power switch Protection lid Pro...

Страница 10: ...ressing to the desired temperature Up and down key is used to adjust the desired temperature value and the desired mode of the temperature LCD display Displays the temperature value and the mode setti...

Страница 11: ...Block Cooler V1 0 9 C Installation of Block Cooler System CB 1 C 1 Package List Item Quantity Block cooler system CB 1 1 48 Well aluminum block 1 Block protection rubber 1 Block extractor 1...

Страница 12: ...with the minimum size of 200mm L x 304mm W x 184mm H 7 87 x 11 97 x 7 24 2 Place the 48 well aluminum block into the block cooler system and place the block protection rubber on top of the block 3 Th...

Страница 13: ...setting the temperature parameter press Run Stop key once to start running the block cooler Note In general the block cooler system will require several minutes to achieve the desired temperature on...

Страница 14: ...y thermometer is not calibrated accordingly to international standard thermometer Some temperature variation may occur with the usage of the different non calibrated thermometers Caution Maximum adjus...

Страница 15: ...e system achieves the desired temperature 4 and stabilized Ready indicator turned on h When the ready indicator turned on ensure the temperature reading on the users calibrated thermometer is as per t...

Страница 16: ...tion Fan is not spinning LED Display do not light up Front panel power on indicator off Check main AC power source Replace a new same configuration fuse and try again Check if power switch malfunction...

Страница 17: ...ntative Overheat Error The block cooler system is overheated during a run temperature of the system exceeded more than 100 C Ensure the block cooler system operating condition temperature 0 26 Contact...

Страница 18: ...lean water Organic solvents or strong detergents may damage the instrument and should not be used 48 Well aluminum block should be protected from organic solvents and detergents and it should be clean...

Страница 19: ...ption 1093001 CB 1 Block Cooler System with block extractor 48 well aluminum block and block protection rubber 100 240VAC 50 60Hz Optional Accessories Item Description 1093101 48 Well aluminum block f...

Страница 20: ...03060 Wealtec Corp 1885 Meadowvale Way Sparks NV 89431 Tel 1 775 351 2066 Fax 1 775 351 2077 http www wealtec com E mail support wealtec com...

Отзывы: