64
3D Phase Measurement Procedure
1
– When the target is in position, select the soft key (or long hold
). The 3D Phase Measurement scan will begin.
Note:
Keep the probe still during the PM surface scan, typically
1 second. Multiple images are taken to complete the scan, and
the probe must be held still during this sequence. Do not move
the probe until the Capturing images... keep probe still caution
disappears.
2
– Select
Measurement
to begin the measurement process.
3
– Select at any time to define a
Measurement Plane
. Once
inserted into an image, all measurements added before or after
the plane was inserted are made with respect to the Measurement
Plane.
Note:
When an application calls for placement of measurement
cursors within an area that does not include measurable
image pixels, inserting a measurement plane may allow the
measurement to be performed.
Click here to learn more about
Placing a Measurement Plane.
4
– Choose desired measurement type.
5
– The Patented Zoom Window automatically opens allowing for
precise positioning of the active cursor. Tap the window’s edges
(or tap on the cursor, then control with the joystick) to adjust the
active cursor position.
Click here to learn about turning the Zoom
Window on or off.
6
– Place cursors on the desired feature. Up to five measurements
may be placed on each image.
Note:
The system will wait up to 1.5 seconds for movement to stop
before starting a 3DPM capture. If movement does not stop, a
message displays and the capture is not attempted. If movement
begins during the capture, the system pauses the capturing
process, waits for movement to stop, and attempts a second
capture. If either attempt allows enough frames to be captured
without movement to provide sufficient surface coverage,
measurement is allowed. If neither attempt is completed without
movement, a message indicating that the capture was affected
by motion is displayed. This may result in more un-measurable
(red) pixels or a slightly higher noise level than would be achieved
with no movement. The system does not allow measurement if the
movement was sufficient to substantially degrade 3D data quality.