![Waygate Technologies Everest Mentor Visual iQ VideoProbe Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/waygate-technologies/everest-mentor-visual-iq-videoprobe/everest-mentor-visual-iq-videoprobe_operating-manual_3913187058.webp)
58
Multi-Segment
Description: The length of a
nonlinear feature or defect.
Cursor Placement:
Place two or more measurement
cursors (up to 24) to create line
segments along the feature. When
three cursors are used, the 3D
angle between the line segments is
displayed along with the total length
result.
When finished, press
twice or
select DONE. The line is finalized.
Note:
With touch-enabled systems
or a connected mouse, additional
cursors can be placed by touching or
clicking on a placed cursor.
Blade Tip Clearance
Description: Automatically
identifies the blade and
shroud, and maps the edge of the
blade, and determines the minimum,
maximum, and average clearance
within the mapped blade edge region.
Cursor Placement:
All cursors are automatically placed
on the tip and shroud. Check the
3D shroud data in the point cloud
view. If large anomalies are present,
adjust the tip position for a view less
perpendicular to the blade.
Otherwise, manually adjust the 3
shroud cursor positions to maximize
green mask coverage and improve
blue rectangle alignment.
If a shorter segment of the blade
tip clearance needs to be further
evaluated, the two bounding cursors,
one at each end of the mapped blade
edge, can be moved to exclude end
portions from the min/max/average
analysis.
Area Depth
Description: Gives a depth
profile through the deepest or
highest point in an area of interest.
Cursor Placement:
Place two cursors to define a line on
the reference surface to one side of
the area of interest and a third cursor
on the opposite side of the area of
interest. The system determines a
second line best matching the surface
curvature along the first line. It then
sweeps depth profiles between the
two lines and identifies the profile that
includes the deepest or highest point.
Note:
Use Run Demo to see all
calculated depth profiles.
Note:
Can be used on flat surfaces or
surfaces with curvature in only one
direction such as pipes.
Area
Description: The surface area
contained within multiple
cursors placed around a feature or
defect.
Cursor Placement:
Place three or more cursors (up to 24)
around the edge of the area you want
to measure.
When finished, press
twice or
select DONE. The area closes.
Note:
Area measurements are
computed assuming a relatively flat
surface. Area measurements on highly
curved or irregular surfaces are less
accurate. The point cloud view shows
the actual area being computed.
Note:
With touch-enabled systems
or a connected mouse, additional
cursors can be placed by touching or
clicking on a placed cursor.