WAVECREST Corporation                                                                               

DTS-2077 Calibration Procedure

(17 April,  1998)

4

3.0

100 MHz Oscillator Check

(3-1)

Connect one SMA cable to the l00 MHz reference clock output at the back of the unit.
Connect the SMA cable to the frequency counter (50 ohm input) and verify the oscillator
output is   100.0 MHz 

±

0.9 Hz.

If necessary, remove the screw on the top surface of the oscillator located in the front half
of the system, and use the non-conductive adjustment tool to change the oscillator
frequency.  (Clockwise adjusts the frequency down.)  Allow frequency to settle for one to
two minutes before readjustment.

(3-2)

When adjusted properly, install screw into oscillator.  Re-install top and bottom covers,
but do not put screws in the top cover at this time.  Allow system to run for one hour
before proceeding.

(3-3)

Repeat the oscillator check and verify the oscillator output is 100 MHz 

±

0.9 Hz before

proceeding.  When ready to proceed, install the screws in the top cover.

4.0

Self Calibration

Note:

The (4) shorting caps and (2) 36” SMA cables provided with the system will be used
in the following steps.

(4-1)

Initiate an internal calibration sequence by pushing the INT CAL button on the front
panel.  The display will indicate the time remaining to accomplish the internal calibration,
counting down on the display.  When the countdown reaches 0 seconds, the display will
change to:  CALIBRATED.  The previous mode and display then returns.

(4-2)

Press the EXT CAL button on the front panel.  The display will direct the user from the
front panel.  Both the DC and AC calibrations should be done at this time.  Follow the
instruction prompts on the front panel display.  DO NOT use shorting caps on the CAL1
or CAL2 output connectors.  Use only on CH1, CH2, ARM1 and ARM2 connectors.
When the instrument completes its calibration sequence, the DTS will display
CALIBRATED.  The previous mode and display then returns.

(4-3)

With the DTS connected to a computer, use the “Virtual Instruments” software or
calibration software to first perform the DTS connect sequence over the GPIB interface,
then perform the STROBE CAL operation.  Follow the instructions on the computer
display.  When the strobe calibration has completed successfully, exit the program and
disconnect the GPIB interface.

This completes the calibration portion of this procedure.  The remaining steps are performance
verification steps to ensure the system is operating properly.

Содержание DTS-2077

Страница 1: ...Calibration Procedure for the DTS 2077 Time Measurement System WAVECREST Corporation...

Страница 2: ...utine expression of that philosophy For this reason any current WAVECREST product may differ in some respect from its published description but will always equal or exceed the original design specific...

Страница 3: ...s 2 Reference Voltage Check 3 3 0 100MHz Oscillator Check 4 4 0 Self Calibration 4 5 0 Reference Voltage Operational Check 5 6 0 Calibration Signal Outputs 5 7 0 AC Performance Checks 6 8 0 Arming and...

Страница 4: ...cable 1 Virtual Instruments or calibration software Notes 1 Use only electronic test equipment with calibrations traceable to NIST standards 2 Refer to the DTS 2077 User s Guide for general operating...

Страница 5: ...ower Supply Checks 2 1 Verify the AC power cord is connected to an appropriate power source 110V 60 Hz or 230V 50 Hz Turn on the main power switch at the rear of the system Then turn on the standby sw...

Страница 6: ...DTS as shown in Figure 1B Verify the voltage is 5VDC 0 05V If adjustment is required both 5VDC modules must be alternately adjusted at VOUT The two 5VDC modules provide load sharing and must be balan...

Страница 7: ...internal calibration sequence by pushing the INT CAL button on the front panel The display will indicate the time remaining to accomplish the internal calibration counting down on the display When th...

Страница 8: ...lower readings on the DTS display It may be necessary to measure and set the DC voltage at the DTS input connectors using a T adapter 6 0 Calibration Signal Outputs 6 1 Using the frequency counter an...

Страница 9: ...s its calibration sequence the DTS will display CALIBRATED The previous mode and display then returns 7 2 Go to Menu 11 see Figure 2 6 in the User s Guide and place the cursor on the CLEAR field Then...

Страница 10: ...the front panel and press FUNC Verify the displayed measurement is 5000 psec 10 psec with JITTER 6 psec 8 3 Set the function generator to output a square wave with a 5 nsec wide duty cycle and a 1V pk...

Страница 11: ...ndby switch on the front panel and the main power switch on the rear of the instrument Remove all cables and unplug the AC power cord from the power source and package the accessories that accompanied...

Страница 12: ...ke Road Edina MN 55439 Yes please include me when you distribute future new versions of the DTS 2077 Calibration Procedure NAME __________________________________________________________ COMPANY _____...

Страница 13: ...adquarters West Coast Office 7275 Bush Lake Road 1735 Technology Drive Suite 400 Edina MN 55439 San Jose CA 95110 612 831 0030 408 436 9000 FAX 612 831 4474 FAX 408 436 9001 Toll Free 1 800 733 7128 1...

Отзывы: