WAVECREST Corporation                                                                               

DTS-2077 Calibration Procedure

(17 April,  1998)

5

5.0

Reference Voltage Operational Check

(5-1)

Using the voltmeter, set the power supply for each value shown in Table 1.  Then remove
the voltmeter and connect the power supply to the DTS CH1.  Press the "FUNC" key to
perform a pulse find operation.  The displayed voltages at Menu #15 for CH1 should be
between the values shown in Table 1.

Table 1. Input Voltage Correlation

Selected Voltage 

Min Reading Max 

Reading

-1.000V

-1.0025V

-0.9975V

-0.250V

-0.2522V

-0.2478V

-0.025V

-0.0271V

-0.0229V

+0.025V

+0.0229V

+0.0271V

+0.250V

+0.2478V

+0.2522V

+1.000V

+0.9975V

+1.0025V

(5-2)

Repeat for CH2.

Note:  Depending on the power supply drive capability, the 50 ohm input impedance of the
DTS may result in slightly lower readings on the DTS display.  It may be necessary to
measure and set the DC voltage at the DTS input connectors using a “T” adapter.

6.0

Calibration Signal Outputs

(6-1)

Using the frequency counter and the oscilloscope and BNC-SMA adapter, if required,
measure the outputs of both the CAL1 and CAL2 connectors.  Verify the outputs are:

2.0Vpp ± 0.2V

200 MHz ± 200 Hz

Note:

It is necessary to terminate both CAL outputs to 50 ohms when measuring outputs.
This can be done by measuring one channel while externally triggering from the
other output channel.

(6-2)

Connect the CAL1 to the CH1 channel, and the CAL2 to the CH2 channel using the SMA
coax cables.

(6-3)

Go to Menu #9 (see Figure 2-6 in the User's Guide) and change MEASURE to ON.  Then
press the BURST button on the front panel.  The display will read:  Subsequent cables
will use this cable arrangement as a reference.  
To verify this, press BURST again.
The value displayed should be less than two picoseconds.

Содержание DTS-2077

Страница 1: ...Calibration Procedure for the DTS 2077 Time Measurement System WAVECREST Corporation...

Страница 2: ...utine expression of that philosophy For this reason any current WAVECREST product may differ in some respect from its published description but will always equal or exceed the original design specific...

Страница 3: ...s 2 Reference Voltage Check 3 3 0 100MHz Oscillator Check 4 4 0 Self Calibration 4 5 0 Reference Voltage Operational Check 5 6 0 Calibration Signal Outputs 5 7 0 AC Performance Checks 6 8 0 Arming and...

Страница 4: ...cable 1 Virtual Instruments or calibration software Notes 1 Use only electronic test equipment with calibrations traceable to NIST standards 2 Refer to the DTS 2077 User s Guide for general operating...

Страница 5: ...ower Supply Checks 2 1 Verify the AC power cord is connected to an appropriate power source 110V 60 Hz or 230V 50 Hz Turn on the main power switch at the rear of the system Then turn on the standby sw...

Страница 6: ...DTS as shown in Figure 1B Verify the voltage is 5VDC 0 05V If adjustment is required both 5VDC modules must be alternately adjusted at VOUT The two 5VDC modules provide load sharing and must be balan...

Страница 7: ...internal calibration sequence by pushing the INT CAL button on the front panel The display will indicate the time remaining to accomplish the internal calibration counting down on the display When th...

Страница 8: ...lower readings on the DTS display It may be necessary to measure and set the DC voltage at the DTS input connectors using a T adapter 6 0 Calibration Signal Outputs 6 1 Using the frequency counter an...

Страница 9: ...s its calibration sequence the DTS will display CALIBRATED The previous mode and display then returns 7 2 Go to Menu 11 see Figure 2 6 in the User s Guide and place the cursor on the CLEAR field Then...

Страница 10: ...the front panel and press FUNC Verify the displayed measurement is 5000 psec 10 psec with JITTER 6 psec 8 3 Set the function generator to output a square wave with a 5 nsec wide duty cycle and a 1V pk...

Страница 11: ...ndby switch on the front panel and the main power switch on the rear of the instrument Remove all cables and unplug the AC power cord from the power source and package the accessories that accompanied...

Страница 12: ...ke Road Edina MN 55439 Yes please include me when you distribute future new versions of the DTS 2077 Calibration Procedure NAME __________________________________________________________ COMPANY _____...

Страница 13: ...adquarters West Coast Office 7275 Bush Lake Road 1735 Technology Drive Suite 400 Edina MN 55439 San Jose CA 95110 612 831 0030 408 436 9000 FAX 612 831 4474 FAX 408 436 9001 Toll Free 1 800 733 7128 1...

Отзывы: