background image

WAVECREST Corporation                                                                               

DTS-2077 Calibration Procedure

(17 April,  1998)

6

(6-4)

Place the delay line between the CH1 input and the SMA cable connected to CAL1.
Verify that CAL2 is connected to the CH2 input.

(6-5)

Press the BURST button on the front panel.  Verify the "electrical length" of the cable
delay line being measured (in picoseconds) is within ±10 psec of the calibrated length.

(6-6)

Set MEASUREMENT to OFF and remove the delay line.

7.0

AC Performance Checks

This step turns on the 200 MHz calibration signal following the cable measurement step.

(7-1)

Verify that CAL1 is connected to CH1 and CAL2 is connected to CH2.  Press the EXT
CAL button on the front panel.  The display will direct the user from the front panel.
Perform only the AC external calibration at this time.  (Do not press “GO” for DC Cal.)
When the instrument completes its calibration sequence, the DTS will display
CALIBRATED.  The previous mode and display then returns.

(7-2)

Go to Menu #11 (see Figure 2-6 in the User's Guide) and place the cursor on the CLEAR
field.  Then press FUNC.  This sets up the DTS in a known defaults state.

(7-3)

Go to Menu #14 and set the Number of Readings (sample size) to l0,000, SETS to l, and
press CONT.  Then go to Menu #1.

(7-4)

With the TRIG field set to 50%-50%, scroll through the various measurement options
shown below and record the measurements.  Verify JITTER (JTrms) measures <6ps
(>2KHz for frequency).  Then connect CAL1 to CH2 and CAL2 to CH1.  Scroll through
the various measurement options shown below and record the measurements.  Average
the two sets of measurements recorded for each measurement option.  The average should
be within the limits shown below.

*Both Channels

TPD++

0.000 psec 

±

 10 psec

TPD--

0.000 psec 

±

 10 psec

TPD+-

2500.000 psec 

±

 10 psec

TPD-+

2500.000 psec 

±

 10 psec

PW+*

2500.000 psec 

±

 10 psec

PW-*

2500.000 psec 

±

 10 psec

PERIOD*

5000.000 psec 

±

 10 psec

FREQ*

200.0 MHz 

±

 1 KHz

Содержание DTS-2077

Страница 1: ...Calibration Procedure for the DTS 2077 Time Measurement System WAVECREST Corporation...

Страница 2: ...utine expression of that philosophy For this reason any current WAVECREST product may differ in some respect from its published description but will always equal or exceed the original design specific...

Страница 3: ...s 2 Reference Voltage Check 3 3 0 100MHz Oscillator Check 4 4 0 Self Calibration 4 5 0 Reference Voltage Operational Check 5 6 0 Calibration Signal Outputs 5 7 0 AC Performance Checks 6 8 0 Arming and...

Страница 4: ...cable 1 Virtual Instruments or calibration software Notes 1 Use only electronic test equipment with calibrations traceable to NIST standards 2 Refer to the DTS 2077 User s Guide for general operating...

Страница 5: ...ower Supply Checks 2 1 Verify the AC power cord is connected to an appropriate power source 110V 60 Hz or 230V 50 Hz Turn on the main power switch at the rear of the system Then turn on the standby sw...

Страница 6: ...DTS as shown in Figure 1B Verify the voltage is 5VDC 0 05V If adjustment is required both 5VDC modules must be alternately adjusted at VOUT The two 5VDC modules provide load sharing and must be balan...

Страница 7: ...internal calibration sequence by pushing the INT CAL button on the front panel The display will indicate the time remaining to accomplish the internal calibration counting down on the display When th...

Страница 8: ...lower readings on the DTS display It may be necessary to measure and set the DC voltage at the DTS input connectors using a T adapter 6 0 Calibration Signal Outputs 6 1 Using the frequency counter an...

Страница 9: ...s its calibration sequence the DTS will display CALIBRATED The previous mode and display then returns 7 2 Go to Menu 11 see Figure 2 6 in the User s Guide and place the cursor on the CLEAR field Then...

Страница 10: ...the front panel and press FUNC Verify the displayed measurement is 5000 psec 10 psec with JITTER 6 psec 8 3 Set the function generator to output a square wave with a 5 nsec wide duty cycle and a 1V pk...

Страница 11: ...ndby switch on the front panel and the main power switch on the rear of the instrument Remove all cables and unplug the AC power cord from the power source and package the accessories that accompanied...

Страница 12: ...ke Road Edina MN 55439 Yes please include me when you distribute future new versions of the DTS 2077 Calibration Procedure NAME __________________________________________________________ COMPANY _____...

Страница 13: ...adquarters West Coast Office 7275 Bush Lake Road 1735 Technology Drive Suite 400 Edina MN 55439 San Jose CA 95110 612 831 0030 408 436 9000 FAX 612 831 4474 FAX 408 436 9001 Toll Free 1 800 733 7128 1...

Отзывы: