Required materials
• Chemical-resistant, powder-free gloves
To remove the probe inlet fitting:
Warning:
To avoid personal contamination with biologically hazardous or toxic compounds,
wear clean, chemical-resistant, powder-free gloves when performing this procedure.
1. Pull the inlet fitting from the end of the probe assembly.
Tip:
Hold the fitting spring retainer in place while the probe inlet fitting is removed to
prevent it from sliding along the probe assembly.
Figure 3–18: Removing the probe inlet fitting
Fitting spring retainer
Fitting spring
Probe inlet fitting
2. Feed the probe assembly through the narrow channel.
3. Slide the probe inlet fitting onto the probe assembly.
4. Push the probe inlet fitting onto the fitting spring retainer, ensuring that the retainer remains
fully attached to the spring.